Monthly cleaning of FacsCalibur with Cytek Fluid Manager
Article Details

Last Updated
22nd of July, 2009

User Opinions (6 votes)
100% thumbs up 0% thumbs down

How would you rate this answer?
Helpful
Not helpful
The Monthly Cleaning Procedure should be performed at regular intervals or when FSC noise is excessive. This procedure cleans the supply lines and flow cell walls with 10% bleach, then rinses the beach out with deionized H2O. It is critical to bypass the sheath filter and to use a dedicated 4L bleach tank for this procedure. Below is a procedure for how to disconnect the 20L Fluid Manager Ultra Precision, perform the cleaning, and reconnect the 20L Fluid Manager Ultra Precision.

See below PDF document for detailed instructions: 
Attachments
Visitor Comments
  1. Comment #1 (Posted by ahmed )
    i am sorry i need help for some happen with me with the FACS cytometry during data acquisition from the instrument i have cuting in the electrical current and then return in the same time now the cytometry work well but no "connect to flow cytometry " option appear in the program i dont know what happen is that burn in the instrument or what?